Table 52. Clinical Management Alerts (continued)
Parameter
Programmable values
Shipped
Reset
Shared (Device Tone and Patient Home Monitor)
Number of Shocks Thresh-
1 ; 2; 3; 4; 5; 6
1
1
All Therapies in a Zone Exhausted for an Episode.
Device Tone
Alert Enable - Urgency
Off ; On-Low; On-High
Off
Off
Patient Home Monitor
Alert Enable
Off ; On
Off
Off
a
Decreasing the OptiVol Threshold will make the device more sensitive to changes in the patient’s thoracic fluid
status. Increasing the OptiVol Threshold could delay or prevent device observation of significant changes in the
patient’s thoracic fluid status.
b
This parameter is displayed only if an associated alert has been enabled.
c
Alerts are programmable and transmittable to a monitor only when Patient Home Monitor is programmed to Yes.
d
Note that VF, VT, and FVT therapies could be delivered during a single episode (from initial detection until
episode termination).
Table 53. Lead/Device Integrity Alerts
Parameter
Programmable values
Shipped Reset
RV Lead…
Device Tone
Low; High
High
—
RV Lead Integrity Enable
On ; Off
On
Off
RV Lead Noise Enable
On ; Off
On
Off
Patient Home Monitor
RV Lead Integrity Enable
On ; Off
On
Off
RV Lead Noise Enable
On ; Off
On
Off
Lead Impedance Out of Range…
Device Tone
Low; High
High
—
A. Pacing Impedance Ena-
ble
On ; Off (Observation only)
On
Off (Observation
only)
RV Pacing Impedance Ena-
ble
On ; Off (Observation only)
On
Off (Observation
only)
RV Defibrillation Impedance
Enable
On ; Off (Observation only)
On
Off (Observation
only)
SVC Defibrillation Impe-
dance Enable
On ; Off (Observation only)
On
Off (Observation
only)
Patient Home Monitor
A. Pacing Impedance Ena-
Off; On
On
Off
Medtronic
PROTECTA™ XT DR D314DRG
472
Clinician Manual