7.10.3.1 How to set the device clock
Select Params icon
⇒ Data Collection Setup…
⇒ Device Date/Time…
7.10.4 Programming the Sleep feature
Select Params icon
⇒ Pacing…
⇒ Additional Features…
⇒ Sleep…
▷ Sleep <On>
▷ Sleep Rate
▷ Bed Time
▷ Wake Time
7.10.5 Evaluation of the Sleep feature
The Ventricular Rate Histogram shows heart rates below the Lower Rate but above the
Sleep Rate for the percentage of time that correlates to the Sleep period. See Section 6.12,
“Using rate histograms to assess heart rates”, page 183 for more information.
The Cardiac Compass Report shows the average ventricular rate during the day and night,
which should indicate that the device is allowing a slower heart rate at night. See Section 6.5,
“Viewing long-term clinical trends with the Cardiac Compass Report”, page 148 for more
information.
7.11 Preventing competitive atrial pacing
An atrial tachycardia may be initiated if an atrial paced event occurs within the vulnerable
period of the atrium. This can happen if the device is pacing at a high rate, if a premature
atrial contraction occurs during an atrial refractory period and is quickly followed by an atrial
pace.
7.11.1 System solution: NCAP
The Non-Competitive Atrial Pacing (NCAP) feature prevents pacing the atrium too soon
after a refractory atrial sense by delaying the scheduled atrial pace.
Medtronic
PROTECTA™ XT DR D314DRG
266
Clinician Manual