Figure 176. Episode EGM for RV Lead Noise Discrimination event
1 “N” annotated upon detection of RV lead noise.
8.8.6.3 CareAlert pop-up window
When the device is interrogated, a CareAlert window notifies you that an alert condition
exists, including the RV Lead Noise alert.
8.8.6.4 Quick Look II observations
Select Data icon
⇒ Quick Look II
Check the Quick Look II Observations list to verify that there is an RV lead noise warning.
For detailed information about viewing and interpreting all of the information available from
the Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
349