programmed therapy. This information is shown on the screen only if AT/AF detection is
programmed to 2 zones.
Treated episodes per cycle length – This counter reports the number of episodes treated
per atrial cycle length and the percentage of successfully terminated episodes per atrial
cycle length.
ATP Sequences – This counter reports the number of atrial ATP sequences delivered and
the number aborted.
9.6 Treating AT/AF with atrial cardioversion
An AT/AF episode is detected when sustained atrial tachycardia occurs. Treatments for
these episodes are intended to interrupt the atrial tachyarrhythmia and restore sinus rhythm.
Atrial ATP therapies may terminate these episodes. High-voltage therapies may terminate
these episodes if ATP therapies are ineffective.
9.6.1 System solution: atrial cardioversion
The device can respond to an AT/AF episode by delivering atrial cardioversion therapy to
the patient’s heart. Cardioversion is intended to terminate the episode by simultaneously
depolarizing the heart tissue and restoring the patient’s normal sinus rhythm.
Atrial cardioversion is delivered automatically when it is scheduled by the device. Optionally,
it can be delivered when the patient uses the patient assistant to request it.
For related information, see the following sections:
●
Section 8.1, “Detecting atrial tachyarrhythmias”, page 295
●
Section 9.4, “Scheduling atrial therapies”, page 389
●
Section 9.7, “Providing patient-activated atrial cardioversion”, page 414
9.6.2 Operation of atrial cardioversion
When an automatic atrial cardioversion (CV) is scheduled to be delivered during an AT/AF
episode, the device charges the high-voltage capacitors to the programmed energy level
and attempts to synchronize the shock to a sensed ventricular event outside the ventricular
vulnerable period. If synchronization cannot occur, the device aborts the therapy.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual