9.7 Providing patient-activated atrial cardioversion
Treatments for AT/AF episodes are intended to interrupt the atrial tachyarrhythmias and
restore sinus rhythm. Although automatic atrial therapies (ATP and cardioversion) are
available to terminate these episodes, it may also be desirable to allow the patient to request
atrial cardioversion.
9.7.1 System solution: patient-activated atrial cardioversion
Patient-activated atrial cardioversion allows the patient to request therapy delivery
according to instructions that you provide in advance. This is intended to give the patient
more control over the setting and timing of the therapy. For a patient-activated therapy, the
patient uses the Model 2696 InCheck Patient Assistant to request delivery of an atrial
cardioversion therapy by the device. The device delivers atrial cardioversion therapy if
certain conditions are met.
For related information, see the following sections:
●
Section 8.1, “Detecting atrial tachyarrhythmias”, page 295
●
Section 9.6, “Treating AT/AF with atrial cardioversion”, page 407
For additional information about the Model 2696 InCheck Patient Assistant, refer to the
manual provided with it.
9.7.2 Operation of patient-activated atrial cardioversion
To request atrial cardioversion, the patient first uses the patient assistant to learn whether
an AT/AF episode is in progress. If an episode is in progress, the patient can request
cardioversion. A patient-activated atrial cardioversion becomes pending when the delivery
conditions are met. When a patient-activated atrial cardioversion is pending, it takes priority
over a scheduled automatic atrial therapy.
When a patient-activated atrial cardioversion is pending, the device charges the
high-voltage capacitors to the programmed energy level and attempts to synchronize the
shock to a sensed ventricular event outside the ventricular vulnerable period. If
synchronization cannot occur, the device aborts the therapy.