Select Params icon
⇒ Pacing…
⇒ Rate Response…
▷ Lower Rate
▷ ADL Rate
▷ Upper Sensor
▷ Rate Profile Optimization
▷ ADL Response
▷ Exertion Response
⇒ Additional Parameters…
▷ Activity Threshold
▷ Activity Acceleration
▷ Activity Deceleration
▷ ADL Setpoint
▷ UR Setpoint
7.4.5 Evaluation of Rate Response
7.4.5.1 Rate Histograms Report
Select Data icon
⇒ Clinical Diagnostics
⇒ Rate Histograms (Report Only)
The Rate Histograms Report provides information about how Rate Response has been
performing since the previous patient session.
In Figure 88, you can see how the histograms changed after Rate Response was
programmed to be more aggressive.
Note that the percentage of atrial pacing has shifted from the lower rates to the higher rates.
Medtronic
PROTECTA™ XT DR D314DRG
236
Clinician Manual