8.7.1 System solution: High Rate Timeout
The High Rate Timeout feature allows the device to deliver therapy for any ventricular
tachyarrhythmia that continues beyond a programmed length of time.
8.7.2 Operation of High Rate Timeout
You can program separate timeout periods for All Zones and for the VF Zone Only.
All Zones – The device starts the programmed High Rate Timeout period when VF, FVT,
or VT detection occurs or is withheld by an SVT discrimination feature. If the tachyarrhythmia
continues beyond the programmed timeout period, the device suspends all SVT
discrimination features and allows the device to deliver therapy.
VF Zone Only – The device starts the programmed High Rate Timeout period when VF or
FVT detection occurs or is withheld by an SVT discrimination feature. If the tachyarrhythmia
continues beyond the programmed timeout period, the device suspends all SVT
discrimination features and allows the device to deliver therapy.
Notes:
●
When a VF Zone Only timeout occurs, and the tachyarrhythmia leaves the VF zone
before the timeout period expires, the timeout period is reset. If the tachyarrhythmia
reenters the VF zone, the timeout starts over from the beginning.
●
If, during either High Rate Timeout period, the device determines that the SVT
discrimination feature no longer applies, detection occurs and therapy is delivered,
regardless of the High Rate Timeout feature.
●
High Rate Timeout is not applied to rates in the VT Monitor zone.
8.7.3 Programming considerations for High Rate Timeout
Programming both High Rate Timeout periods – If both timeout periods are
programmed, the shorter timeout is applied first. The VF Zone Only timeout would normally
be programmed with the shorter duration.
High Rate Timeout and inappropriate therapies – After the programmed High Rate
Timeout period elapses and the SVT discrimination features are suspended, it is possible
that the device may detect a ventricular tachyarrhythmia that is actually a conducted SVT.
If so, inappropriate tachyarrhythmia therapy may be delivered.