![Medtronic PROTECTA XT DR D314DRG Скачать руководство пользователя страница 209](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148209.webp)
Repositioning the atrial lead – You may need to reposition or replace the atrial sensing
lead if reprogramming the atrial sensing threshold, set by reprogramming the A. Sensitivity
parameter, does not provide reliable atrial sensing during AT/AF episodes and sinus rhythm.
Absolute PVAB – PVAB Method cannot be set to Absolute when the programmed pacing
mode is ODO, AAI, or AAIR.
Upper rates and refractory periods – A combination of high Upper Sensor Rate, high
Upper Tracking Rate, and a long refractory period may result in competitive atrial pacing.
For more information, see Section 7.11, “Preventing competitive atrial pacing”, page 266.
Low sensing threshold – If you set a sensitivity parameter to its most sensitive value, the
device is more susceptible to electromagnetic interference (EMI), cross-chamber sensing,
and oversensing.
Recommended ventricular sensing threshold – Setting RV Sensitivity to 0.3 mV is
recommended to maximize the probability of detecting VF and to limit the possibility of
oversensing and cross-chamber sensing.
Recommended atrial sensing threshold – Setting A. Sensitivity to 0.3 mV is
recommended to optimize the effectiveness of atrial detection and pacing operations while
limiting the possibility of oversensing and cross-chamber sensing.
Testing sensitivity after reprogramming – If you change the ventricular sensing
threshold or the ventricular sensing polarity, evaluate for proper sensing. If appropriate, test
for proper detection by inducing VF and allowing the device to automatically detect and treat
the tachyarrhythmia.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
209
Содержание PROTECTA XT DR D314DRG
Страница 2: ......
Страница 9: ...Index 488 Medtronic PROTECTA XT DR D314DRG Clinician Manual 9 ...
Страница 506: ......
Страница 507: ......