Transvenous leads – If you use a subclavian approach to implant a transvenous lead,
position the lead laterally to avoid pinching the lead body between the clavicle and the first
rib.
Epicardial leads – A variety of surgical approaches can be used to implant epicardial leads,
including a limited left thoracotomy or median sternotomy. A typical placement may use an
anterior right ventricular patch as the RVcoil (HVB) and a posterolateral left ventricular patch
as the SVC (HVX).
5.3 Testing the lead system
After the leads are implanted, test the lead system to verify that the sensing and pacing
values are acceptable.
5.3.1 Considerations for testing the lead system
Bipolar leads – When measuring sensing and pacing values, measure between the tip
(cathode) and ring or coil (anode) of each bipolar pacing/sensing lead.
Lead positioning – Final lead positioning should attempt to optimize pacing threshold,
sensing, and defibrillation threshold if appropriate.
5.3.2 How to verify and save the sensing and pacing values
Medtronic recommends that you use a Model 2290 Analyzer to perform sensing and pacing
measurements. When the analyzer and the device sessions are running concurrently, you
can export the saved lead measurements from the analyzer session into the patient
information parameters in the device session. Refer to the analyzer technical manual for
detailed procedures about performing the lead measurements.
Note: If you perform the lead measurements using an implant support instrument other than
a Model 2290 Analyzer, you must manually enter the measurements in the device session.
Note: The intracardiac EGM telemetered from the device cannot be used to directly assess
sensing.