![Medtronic PROTECTA XT DR D314DRG Скачать руководство пользователя страница 271](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148271.webp)
7.14 Reducing inappropriate ventricular inhibition using
VSP
In a dual chamber pacing system with atrial and ventricular pacing and ventricular sensing,
the device may sense an atrial pacing pulse on the ventricular channel and inhibit ventricular
pacing (crosstalk). When inhibition of ventricular pacing occurs, the device may not provide
full ventricular support.
7.14.1 System solution: VSP
Ventricular Safety Pacing (VSP) detects crosstalk by monitoring for nonphysiologic
ventricular sensed events and responds by pacing the ventricle.
7.14.2 Operation of VSP
VSP is available when the pacing mode is programmed to DDDR, DDD, DDIR, DDI, or an
MVP mode (AAIR<=>DDDR or AAI<=>DDD). VSP functions when the device is operating
in the DDDR, DDD, DDIR or DDI mode.
The device uses a 110 ms VSP window to monitor for ventricular senses that occur too soon
after an atrial pacing pulse. Ventricular senses in the VSP window are classified as
nonphysiologic and are likely due to crosstalk. If a ventricular sensed event occurs within
the VSP window, the device delivers a VSP pulse at the end of the VSP window.
If the sensed event is a result of crosstalk, the backup pacing pulse provides ventricular
support. If the sensed event is a ventricular depolarization, the backup pacing pulse occurs
soon enough to fall in the absolute refractory period of the ventricle to avoid pacing on the
T-wave.
Figure 113. VSP pulse delivered at the end of the VSP window (110 ms)
(VSP window)
110 ms
AV
AP
VP
VP
VP
VS
AP
AP
AV
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
271
Содержание PROTECTA XT DR D314DRG
Страница 2: ......
Страница 9: ...Index 488 Medtronic PROTECTA XT DR D314DRG Clinician Manual 9 ...
Страница 506: ......
Страница 507: ......