page 321. For more information about the Wavelet feature, see Section 8.4, “Discriminating
VT/VF from SVT using Wavelet”, page 326.
AT/AF detection – If AT/AF Detection is programmed to On, VF Initial Beats to Detect must
be less than or equal to 30/40; Monitored VT Beats to Detect must be less than or equal to
36; and VT Initial Beats to Detect must be less than or equal to 36.
VF detection backup – To ensure VF detection backup during VT, FVT, and VT Monitor
episodes, if VT, FVT, or VT Monitor is programmed to On, VF Detection must also be
programmed to On.
Monitored VT Beats to Detect – The Monitored VT Beats to Detect value must be greater
than the VF and VT Initial Beats to Detect.
8.2.4 Programming ventricular tachyarrhythmia detection
Select Params icon
⇒ Detection (V.)…
▷ VF <On>
▷ VF Initial
▷ VF Redetect
▷ VF V. Interval (Rate)
▷ FVT <via VF, via VT>
▷ FVT V. Interval (Rate)
▷ VT <On>
▷ VT Initial
▷ VT Redetect
▷ VT V. Interval (Rate)
▷ Monitor <Monitor>
▷ Monitor Initial
▷ Monitor V. Interval (Rate)
8.2.5 Evaluation of VT/VF detection
8.2.5.1 Quick Look II screen
Select Data icon
⇒ Quick Look II
The Quick Look II screen shows the number of monitored and treated VT, FVT, and VF
episodes since the last session.
Medtronic
PROTECTA™ XT DR D314DRG
316
Clinician Manual