6.13.1.5 Atrial Lead Position Check results
The device can be programmed to automatically disable atrial tachyarrhythmia therapies if
the daily Atrial Lead Position Check identifies a potential problem with the lead position. The
Battery and Lead Measurements screen displays the result of the most recent Atrial Lead
Position Check. For more information about the Atrial Lead Position Check, see Section 9.4,
“Scheduling atrial therapies”, page 389.
6.13.1.6 Lead impedance and sensing amplitude measurements
The Battery and Lead Measurements screen displays recent lead impedance and sensing
amplitude measurements. For lead impedance measurements, the screen displays the
most recent manually-performed measurements or the most recent daily automatic
measurements. For sensing amplitude measurements, the screen displays the most recent
daily automatic measurements. Measurements performed with the manual Sensing Test
are not displayed on the Battery and Lead Measurements screen. For more information
about performing manual lead impedance measurements, see Section 10.4, “Measuring
lead impedance”, page 432. For more information about performing manual sensing
amplitude measurements, see Section 10.5, “Performing a Sensing Test”, page 432.
You can compare the most recent measurements to the trends of daily automatic
measurements by selecting the Lead Impedance [>>] button or Sensing [>>] button to view
the Lead Trends screen.
6.13.2 Viewing lead impedance trends
Every day at 3:00 AM, the device automatically measures the lead impedance on each
implanted lead using subthreshold electrical pulses. These pulses are synchronized to
sensed or paced events and do not capture the heart.
The daily automatic lead impedance measurements are displayed on the Lead Trends
screen, which plots the data as a graph. The graph displays up to 15 of the most recent
measurements and up to 80 weekly summary measurements (showing minimum,
maximum, and average values for each week). Significant or sudden changes in lead
impedance may indicate a problem with the lead.
If the device is unable to perform automatic lead impedance measurements, gaps are
present in the trend graph.
Note: The RV Defib impedance is measured and displayed for the currently programmed
defibrillation pathway only. Reprogramming the Active Can/SVC Coil parameter changes
the electrodes included in the defibrillation pathway, and affects which of the collected
measurements are displayed in the trend graph.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
189