9.9.2.1 Interval management
The Automatic Capacitor Formation feature manages the interval between capacitor
formations. Each time the capacitors are formed, the device schedules the next formation
based on the currently programmed Minimum Auto Cap Formation Interval.
Restarting the interval – Each time a full capacitor formation occurs, the feature restarts
the interval. The restart prevents capacitor formations from occurring too frequently, thereby
preventing unnecessary battery drain.
Figure 216. Interval restart
1
2
3
4
0
6
7
8
9
5
11
12
10
interval
months
1 At month 4, a full energy defibrillation therapy was aborted and allowed to dissipate for at least
10 min. The resulting full capacitor formation restarts the Minimum Auto Cap Formation Interval.
Extending the interval – When a full energy charge is delivered or dumped prematurely
(before 10 min elapses), the interval is automatically extended 2 months due to the resulting
partial capacitor formation. Therefore, the device extends the capacitor formation interval
with each partial formation, as illustrated in Figure 217. The total of any subsequent
extensions will not exceed the programmed Minimum Auto Cap Formation Interval.
Figure 217. Extended interval
1
2
3
4
0
6
7
8
9
5
11
12
10
interval
months
1 Full energy charge for a defibrillation therapy was delivered at month 3 of a 6-month interval.
2 The interval extends 2 more months, from month 6 to month 8. The subsequent interval from
month 3 to month 8 is 5 months.
Medtronic
PROTECTA™ XT DR D314DRG
422
Clinician Manual