7.1.8.3 Viewing P-wave and R-wave amplitude trends
Select Data icon
⇒ Device/Lead Diagnostics
⇒ P/R Wave Amplitude Trends
Figure 69. R-wave Amplitude trend
Every day at 2:15 AM, the device begins to measure the amplitude of intrinsic sensed
events. The device attempts to measure the amplitude of 9 normal intrinsic sensed events,
and then records the median value from those events. If the device has not collected 9
amplitude measurements by midnight, no measurement is recorded. The sensing amplitude
trend graph shows a gap for that day.
7.2 Providing pacing therapies
Patients have a variety of conditions for which pacing therapy may be indicated. These
conditions include cardiac asystole, chronic AT/AF, loss of atrioventricular (AV) synchrony,
or poor ventricular function due to heart failure.