4.8 Saving and retrieving a set of parameter values
Custom sets of parameter values can be saved on the programmer hard drive and retrieved
either in the current patient session or in subsequent patient sessions. This allows you to
save and quickly access a custom set of parameter values for a particular clinical situation.
For example, you may want to save a set of parameter values for an initial implant setting,
for a specific disease state, or for situations in which you need to repeatedly program a
particular set of parameters.
The [Save…] button opens a window where you can assign a name to the set of parameter
values presently displayed by the Parameters screen. A saved parameters set can include
both programmed and pending values. The [Get…] button opens the Get Parameter Set
window to retrieve a Medtronic Nominals parameter set, an Initial Interrogation parameter
set, or a custom parameter set.
4.8.1 How to save a set of parameter values
1. Select the Params icon. Make the desired parameter selections.
2. Select [Save…] to open the Parameter Set Name window.
3. Type a name for the parameter set, and select either [OK] or [ENTER].
4. If a parameter set exists with that name, you either need to confirm that you want to
replace the existing set with a new set, or you need to change the name of the new set
of parameters.