Notes:
●
The time annotations displayed on the report are based on the device clock.
●
You cannot manually clear the Cardiac Compass trend data.
6.5.1 How to print the Cardiac Compass Report
You can print the Cardiac Compass Report starting from either the Data or Reports icon.
Select Data icon
⇒ Clinical Diagnostics
⇒ Cardiac Compass Trends (Report Only)
Select Reports icon
⇒ Available Reports…
⇒ Cardiac Compass Trends
6.5.2 Information provided by the Cardiac Compass Report
The Cardiac Compass Report shows events that have occurred during the reporting period.
It also provides trend graphs that can help you to assess the frequency of VT/VF
arrhythmias, AT/AF arrhythmias, pacing and rate response, and information related to heart
failure.
6.5.2.1 Event information
Figure 34. Event annotations
1
2
A
V
3
1 Current session indicator
2 Last session indicator
3 High-voltage therapy indicator
Medtronic