Magnetic mattress pads or pillows – Items containing magnets can interfere with the
normal operation of a cardiac device if they are within 15 cm (6 in) of the device. Avoid
using magnetic mattress pads or pillows because they cannot easily be kept away from
the device.
●
Portable electric generators up to 20 kW – Keep a cardiac device at least 30 cm (12 in)
away from portable electric generators.
●
UPS (uninterruptible power source) up to 200 A – Keep a cardiac device at least 30 cm
(12 in) away from a UPS. If the UPS is operating by battery source, keep a cardiac device
at least 45 cm (18 in) away.
Home power tools – Most home power tools should not affect cardiac devices. Consider
the following common-sense guidelines:
●
Keep all equipment in good working order to avoid electrical shock.
●
Be certain that plug-in tools are properly grounded (or double insulated). Using a ground
fault interrupter outlet is a good safety measure (this inexpensive device prevents a
sustained electrical shock).
Some home power tools could affect cardiac device operation. Consider the following
guidelines to reduce the possibility of interference:
●
Electric yard and hand-held power tools (plug-in and cordless) – Keep a cardiac device
at least 15 cm (6 in) away from such tools.
●
Soldering guns and demagnetizers – Keep a cardiac device at least 30 cm (12 in) away
from these tools.
●
Gasoline-powered tools and gasoline-powered yard equipment – Keep a cardiac
device at least 30 cm (12 in) away from components of the ignition system. Turn off the
motor before making adjustments.
●
Car engine repair – Turn off car engines before making any adjustments. When the
engine is running, keep a cardiac device at least 30 cm (12 in) away from components
of the ignition system.
Industrial equipment – After recovering from implant surgery, you likely will be able to
return to work, school, or daily routine. However, if you will be using or working near
high-voltage equipment, sources of high electrical current, magnetic fields, or other EMI
sources that may affect device operation, consult with your doctor. You may need to avoid
using, or working near, the following types of industrial equipment:
●
Electric furnaces used in the manufacturing of steel
●
Induction heating equipment and induction furnaces, such as kilns
●
Industrial magnets or large magnets, such as those used in surface grinding and
electromagnetic cranes