Testing for cross-stimulation – At implant, and regularly when atrial ATP therapy is
enabled, conduct testing at the programmed atrial ATP output settings to ensure that
ventricular capture does not occur. This is particularly important when the lead is placed in
the inferior atrium.
Twiddler’s syndrome – Twiddler’s syndrome, the tendency of some patients to manipulate
their device after implant, may cause the pacing rate to increase temporarily if the device is
programmed to a rate-responsive mode.
2.5.1 Pacemaker-dependent patients
Ventricular Safety Pacing – Always program Ventricular Safety Pacing (VSP) to On for
pacemaker-dependent patients. Ventricular Safety Pacing prevents ventricular asystole
due to inappropriate inhibition of ventricular pacing caused by oversensing in the ventricle.
ODO pacing mode – Pacing is disabled under ODO pacing mode. Do not program the
ODO mode for pacemaker-dependent patients. Instead, use the Underlying Rhythm Test
to provide a brief period without pacing support.
Underlying Rhythm Test – Use caution when using the Underlying Rhythm Test to inhibit
pacing. The patient is without pacing support when pacing is inhibited.
2.6 Warnings, precautions, and guidance for clinicians
performing medical procedures on cardiac device patients
This section is intended for physicians and other health care professionals who perform
medical procedures on patients with Medtronic implanted cardiac device systems and who
consult with the patients’ cardiologists. This section provides warnings, precautions, and
guidance related to medical therapies and diagnostic procedures that may cause serious
injury to a patient, interfere with a Medtronic implanted cardiac device system, or
permanently damage the system.
Note: Some common medical procedures that pose no risk are also listed in this section.
For additional guidance on medical procedures not addressed in this section, customers
can contact the following resources:
●
Customers in the United States can contact either of the following telephone numbers:
for pacemakers, contact Medtronic Technical Services at +1 800 505 4636; for ICDs,
contact Medtronic Technical Services at +1 800 723 4636. You may also submit
questions to [email protected] or your Medtronic representative.
●
Customers outside of the United States can contact a Medtronic representative.