Figure 205. Overview of atrial ATP sequence delivery
ATP sequence
scheduled
Monitor for atrial
tachyarrhythmia
events
Atrial tachyarrhythmia
detected
Deliver ATP
sequence
9.5.2.2 Atrial ATP therapy pacing rate and output
Minimum limit for atrial ATP pacing interval – The Burst+ and Ramp pacing intervals
are based on programmed percentages of the atrial tachycardia cycle length, which is
calculated as the median of the last 12 atrial intervals prior to therapy delivery. The median
atrial tachycardia cycle length can vary from one sequence in a therapy to the next, and the
ATP pacing intervals vary accordingly.
The programmable A-A Minimum ATP Interval parameter limits the pacing intervals at which
the Burst+ and Ramp pacing pulses are delivered. If some calculated intervals are shorter
than the programmed A-A Minimum ATP Interval, the pulses are delivered at the A-A
Minimum ATP Interval.
If the median of the last 12 A-A intervals is shorter than the programmed A-A Minimum ATP
Interval, the device does not deliver Burst+ or Ramp therapies until the atrial rate slows.
Note: There is no minimum interval limit for the atrial 50 Hz Burst therapy.
Pacing output for ATP therapies – The A. Pacing Amplitude and A. Pacing Pulse Width
parameter values are the same for all atrial ATP therapies, but they are programmed
separately from the pacing amplitude and pulse width for bradycardia pacing pulses.
Medtronic
PROTECTA™ XT DR D314DRG
398
Clinician Manual