– If appropriate for the patient, program the device to an asynchronous pacing mode
(for example, DOO). After completing radiotherapy treatment, restore device
parameters.
●
Device damage – Exposing the device to high doses of direct or scattered radiation from
any source that results in an accumulated dose greater than 500 cGy may damage the
device. Damage may not be immediately apparent. If a patient requires radiation
therapy from any source, do not expose the device to radiation that exceeds an
accumulated dose of 500 cGy. To limit device exposure, use appropriate shielding or
other measures. For patients who are undergoing multiple radiation treatments,
consider the accumulated dose to the device from previous exposures.
Note: Normally, the accumulated dose from diagnostic radiology is not sufficient to
damage the device. See “Diagnostic radiology” for precautions.
●
Device operational errors – Exposing the device to scattered neutrons may cause
electrical reset of the device, errors in device functionality, errors in diagnostic data, or
loss of diagnostic data. To help reduce the chance of electrical reset due to neutron
exposure, deliver radiotherapy treatment by using photon beam energies less than or
equal to 10 MV. The use of conventional x-ray shielding during radiotherapy does not
protect the device from the effects of neutrons. If photon beam energies exceed 10 MV,
Medtronic recommends interrogating the device immediately after radiotherapy
treatment. An electrical reset requires reprogramming of device parameters. Electron
beam treatments that do not produce neutrons do not cause electrical reset of the
device.
Stereotaxis – Stereotaxis is a catheter navigation platform that allows clinicians to steer
catheter-based diagnostic and therapeutic devices throughout the body by using magnetic
navigation. During a stereotaxis procedure, the magnetic field may activate the magnet
detection sensor in the implanted device, which suspends tachyarrhythmia detection. The
device resumes normal programmed operation after the procedure.
Transcutaneous electrical nerve stimulation (TENS) – TENS (including neuromuscular
electrical stimulation or NMES) is a pain control technique that uses electrical impulses
passed through the skin to stimulate nerves. A TENS device is not recommended for
in-home use by cardiac device patients due to a potential for oversensing, inappropriate
therapy, or inhibition of pacing. If a TENS device is determined to be medically necessary,
contact a Medtronic representative for more information.
Transurethral needle ablation (Medtronic TUNA therapy) – Transurethral needle
ablation is a surgical procedure used for benign prostatic hyperplasia (BPH) in which
precisely focused, conducted radio frequency energy is used to ablate prostate tissue.
Patients with implanted cardiac devices may conditionally undergo procedures that use the
Medtronic TUNA system. To avoid affecting cardiac device function when performing the
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
33