6.3.4 Programming considerations for Medtronic CareAlert feature
Transmitting alerts – The ability to transmit Medtronic alerts to the CareLink Network is
programmable only when the Patient Home Monitor field on the programmer screen is
programmed to Yes.
Repetitive alerts – If a programmable alert is triggered so often that it loses its clinical value,
you may want to consider adjusting the alert threshold, programming the device to improve
therapy effectiveness, or turning off the alert.
6.3.5 Programming Alerts
Select Params icon
⇒ Alert…
⇒ Clinical Management Alerts…
Select Params icon
⇒ Alert…
⇒ Lead/Device Integrity Alerts…
6.3.6 Evaluation of alert events
The device stores alert events in the Medtronic CareAlert Events log. The programmer
screen refers to alert events as Alert Events and OptiVol Events. For each alert event, a log
entry includes the date and time of the alert, a description of the event, and the measurement
or information that caused the event. Up to 15 Alert Events and the last 7 OptiVol Events
are stored. Alert data is cleared only when you clear all device data from the Data Collection
Setup screen.
Caution: Verify lead integrity when evaluating OptiVol Events. Loss of RVcoil integrity due
to lead fracture or insulation defect may adversely affect OptiVol Events.
Select Data icon
⇒ Alert Events
▷ Alert Events
▷ OptiVol Events
Medtronic
PROTECTA™ XT DR D314DRG
140
Clinician Manual