7.17.1 System solution: atrial intervention pacing features
The system provides overdrive pacing techniques that are designed to counteract potential
atrial tachyarrhythmia initiating mechanisms.
Atrial Rate Stabilization (ARS) adapts the pacing rate in response to a PAC to avoid long
sinus pauses following short atrial intervals (short-long-short sequences that may cause the
onset of some atrial tachycardias).
Atrial Preference Pacing (APP) is designed to maintain a consistent activation sequence by
providing continuous pacing that is closely matched to the intrinsic sinus rate.
Post Mode Switch Overdrive Pacing (PMOP) works with the Mode Switch feature to deliver
overdrive atrial pacing during the vulnerable phase following an AT/AF episode termination.
7.17.2 Operation of ARS
Atrial Rate Stabilization (ARS) is available when the pacing mode is programmed to DDDR,
DDD, AAIR, AAI, or MVP (AAIR<=>DDDR or AAI<=>DDD) mode, and functions when the
device is operating in one of these modes.
Figure 118. Atrial Rate Stabilization (ARS)
A
tri
al
ra
te
min
-1
(bp
m
)
Time (s)
Atrial Rate Stabilization
PAC
PAC
50
60
70
80
90
100
110
120
130
140
0
5
10
15
20
Scheduled pace
Paced beats
Intrinsic beats
Atrial Rate Stabilization pacing
Atrial rate