7.17.3 Programming considerations for ARS
Non-Competitive Atrial Pacing (NCAP) – The NCAP feature may delay an atrial pacing
pulse that results from Atrial Rate Stabilization.
Programming constraints – To ensure reliable tachyarrhythmia detection, the system
regulates the values that you can select for the Maximum Rate, Upper Rate, AT/AF
Detection Interval, VT/VF Detection Interval, and Ventricular Monitor interval.
7.17.4 Programming ARS
Select Params icon
⇒ Pacing…
⇒ Arrhythmia/Post Shock…
▷ A. Rate Stabilization <On>
⇒ Additional A Settings…
▷ Maximum Rate
▷ Interval Percentage Increment
7.17.5 Operation of APP
Atrial Preference Pacing (APP) is available when the pacing mode is programmed to DDDR,
DDD, AAIR, AAI, or MVP (AAIR<=>DDDR or AAI<=>DDD) mode, and functions when the
device is operating in one of these modes.
Medtronic
PROTECTA™ XT DR D314DRG
280
Clinician Manual