6.3.3.2 Selecting a patient alert time
The system allows you to select the time of day that a patient alert sounds. The alert tone
continues to sound each day at the selected time until the device is interrogated. Select a
time when the patient or a companion is most likely to hear the tone. The following patient
factors may influence your selection of the alert time:
●
when the patient will be in a predictably quiet setting
●
the patient’s daily schedule; for example, medication routines that may affect alertness
●
the patient’s hearing acuity
●
the presence or absence of companions who might also hear the tones
Note: Some patient alerts will also sound at times other than the programmed alert time.
See Section 6.3.1.1 and Section 6.3.1.2 for exceptions.
If the conditions that trigger an alert are intermittent, the alert event may not actually be
present when the alert tone sounds. Also, the alert time is based on the internal device clock.
It does not adjust for time zone changes.
6.3.3.3 Programming a patient alert time
Select Params icon
⇒ Alert…
⇒ Alert Time… | Device Tone
▷ Alert Time
Tones for some system alerts are synchronized to the time you program tones for
clinician-defined alerts. However, if you program all clinician-defined alerts to Off, the Alert
Time field is not shown. In this situation, perform the following steps to program the Alert
Time for these system alert tones:
1. Program one of the programmable alerts to On in order to display the Alert Time field.
2. Select an alert time to apply to system alerts.
3. Reprogram the alert to Off.
6.3.3.4 Instructing the patient
It is important that patients understand that they may hear alert tones emitted from implanted
devices. They must know what to do when an alert sounds.