6.3 Automatic alerts and notification of clinical
management and system performance events
Important clinical management and system performance events may occur between
scheduled patient sessions. These events may relate to clinical management data stored
in device memory or to inappropriate programmed settings or system issues that should be
investigated. The early detection and notification of these events, should they occur,
enables you to intervene promptly with appropriate care for your patient.
6.3.1 System solution: Medtronic CareAlert feature
The device continuously monitors for a specified set of clinical management and system
performance events that may occur between scheduled follow-up sessions. If the device
detects that such an event has occurred, and if alerting parameters are turned on, it
responds in the following ways:
●
wireless signal and network transmission of event information
Medtronic CareAlert Monitoring continuously monitors for alert events. Should an event
occur, CareAlert Monitoring sends a wireless alert signal to your patient’s Medtronic
CareLink Monitor Model 2490C. Upon receiving a signal, the monitor transmits the alert
data to the Medtronic CareLink Network, if programmed to do so.
●
notification of alert event
Depending on the severity of the alert condition, you can set up Medtronic CareAlert
Notifications through the CareLink Network to hold the alert for routine review at your
office via email or the CareLink website, or to immediately notify you via voice message,
text message, pager, or live call.
●
patient alert
The device emits one of two tones to alert your patient, depending on the type and
urgency of the event that has occurred. The patient then responds according to your
instructions.
6.3.1.1 Clinician-defined alerts (programmable)
Clinical Management Alerts
Possible Fluid Accumulation This alert indicates that the OptiVol 2.0 Fluid Index has met or excee-
ded the programmed OptiVol Threshold. Note that the OptiVol Alert
Enable parameter is permanently programmed to “Off (Observation
only)”, but the OptiVol Threshold parameter is programmable. If the
OptiVol 2.0 Fluid Index has exceeded the programmed threshold,
the fluid accumulation event is listed on the OptiVol Events screen
and is provided as an observation on the Quick Look II screen.