7.8.3 Programming considerations for Rate Drop Response
Symptoms during sleep – During sleep, a patient’s sinus rate may fall below the
programmed Lower Rate, thereby triggering intervention pacing at an inappropriate time.
There are two ways to address this problem. First, you can turn off Low Rate Detection.
Second, you can turn on the Sleep feature. The Sleep feature replaces the programmed
Lower Rate with a slower pacing rate during the time of day the patient normally sleeps. For
more information about the Sleep feature, see Section 7.10, “Providing a slower pacing rate
during periods of sleep”, page 264.
Features that adjust pacing rate – Features that adjust the pacing rate, such as Atrial
Rate Stabilization and Ventricular Rate Stabilization, are unavailable when Rate Drop
Response is programmed on.
7.8.4 Programming Rate Drop Response
Select Params icon
⇒ Pacing…
⇒ Additional Features…
⇒ Rate Drop Response…
▷ Rate Drop Response <On>
▷ Mode
▷ Lower Rate
▷ Detection Type
▷ Drop Size
▷ Drop Rate
▷ Detection Window
▷ Detection Beats
▷ Intervention Rate
▷ Intervention Duration
7.8.5 Evaluation of Rate Drop Response
The Rate Drop Response Episodes screen provides beat-to-beat data that is useful for
analyzing Rate Drop Response episodes and the events that lead up to them. It also
provides information that may help you select appropriate Rate Drop Response detection
parameters.
Medtronic
PROTECTA™ XT DR D314DRG
260
Clinician Manual