7.9.4 Programming Rate Hysteresis
Select Params icon
⇒ Pacing…
⇒ Additional Features…
▷ Rate Hysteresis
7.9.5 Evaluation of Rate Hysteresis
The Ventricular Rate Histogram indicates when the device has allowed the patient’s intrinsic
heart rhythm to prevail at rates lower than the Lower Rate.
7.9.5.1 Viewing the Ventricular Rate Histogram Report
Select Data icon
⇒ Clinical Diagnostics
⇒ Rate Histograms (Report Only)
7.10 Providing a slower pacing rate during periods of sleep
Some patients have difficulty sleeping when they are paced at a rate that is intended for
times when they are normally awake.
7.10.1 System solution: Sleep feature
The Sleep feature replaces the programmed Lower Rate with a slower pacing rate during
the time of day that the patient normally sleeps.
Medtronic
PROTECTA™ XT DR D314DRG
264