If either the Device Tone or the Patient Home Monitor alert parameter for the RV Lead
Integrity Alert is programmed to On, the device responds to a potential RV lead fracture with
an alert tone, a Medtronic CareAlert Notification, and automatic adjustments to
tachyarrhythmia detection. When at least 1 High Rate-NS episode occurs, or if the alert
criteria are met, the device automatically adjusts EGM storage operation.
If both the Device Tone and the Patient Home Monitor alert parameters for the RV Lead
Integrity Alert are programmed to Off, the device does not trigger a Medtronic CareAlert
Notification or an alert tone, and it does not adjust tachyarrhythmia detection parameters.
However, it records a Quick Look II observation for the lead warning, and it makes
adjustments to EGM storage.
6.4.2.1 RV Lead Integrity Alert tones and CareAlert Notifications
When the criteria for the RV Lead Integrity Alert are met, the device immediately sounds an
alert tone. The tone sounds again every 4 hours beginning at the next scheduled 4-hour
time interval (12:00 AM, 4:00 AM, 8:00 AM…). The tone also sounds at the programmed
Alert Time and when a magnet is placed over the device. The tone continues to sound until
the device is interrogated by a programmer. If the Patient Home Monitor alert for the RV
Lead Integrity Alert is programmed to On, the device also attempts to send a wireless
transmission to the Medtronic CareLink Monitor.
6.4.2.2 Automatic adjustments to tachyarrhythmia detection
When the RV Lead Integrity Alert is triggered, the device automatically programs the VF
Initial Beats to Detect parameter to 30/40 (if it was less). If necessary, the device
automatically adjusts the Monitored VT Beats to Detect parameter and sets AT/AF
Detection to Monitor.
In addition, after the RV Lead Integrity Alert has been triggered, the device automatically
suspends tachyarrhythmia detection when it is interrogated by a programmer, CareLink
Monitor, or CardioSight monitor, using nonwireless telemetry. The device suspends
detection to prevent inappropriate therapy delivery during the telemetry session when a
lead issue is suspected.
During a CareLink or CardioSight monitor session, tachyarrhythmia detection resumes
automatically when the interrogation is complete. During a programmer session,
tachyarrhythmia detection can be resumed by selecting [Resume] at the top of the
programmer screen, by removing the programming head, or by ending the session and
removing the programming head.
Note: This suspension behavior occurs for all nonwireless sessions until 3 days have
passed since the alert was cleared by interrogating the device with a programmer.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
143