Figure 132. AT/AF onset and AT/AF detection
M
S
A
S
A
S
A EGM
Marker Channel
V EGM
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
S
T
Dl
T
Dl
T
Dl
T
Dl
T
Dl
V
P
V
P
V
P
V
P
V
P
V
P
V
P
V
P
V
P
V
P
200 ms
1 The MS marker indicates that Mode Switch has taken place. This marker appears only if Mode
Switch has been programmed to On.
2 The TD marker indicates that AT/AF episode detection has taken place.
Notes:
●
When AT/AF detection occurs, the system creates an episode record marking the
AT/AF onset and detection points. If onset is reached but detection never occurs, there
will not be an episode record for that instance of AT/AF.
●
When there are at least 2 atrial events in a ventricular interval, the device analyzes A:V
pattern information to determine if one of the atrial events is actually a far-field R-wave.
Far-field R-waves are not counted toward AT/AF detection.
●
VT/VF detection takes priority over AT/AF detection. When VT/VF is detected, any
ongoing AT/AF detection process is postponed until after the VT/VF episode
terminates.
8.1.2.3 Classifying atrial tachyarrhythmia episodes for treatment
The system uses programmable “detection zones” to classify atrial tachyarrhythmias for
treatment. You can program 1 detection zone (AT/AF) or 2 detection zones (AT/AF and Fast
AT/AF). Use 1 zone if the patient exhibits one clinical atrial tachyarrhythmia. Use 2 zones if
the patient exhibits two distinct clinical atrial tachyarrhythmias and you want to treat each
tachyarrhythmia with a unique set of therapies.
Figure 133. AT/AF and Fast AT/AF detection parameters
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
297