Figure 44. Example of Heart Failure Management Report
The OptiVol 2.0 Fluid Index is a plot of the accumulation of differences, adjusted for
individual patient variation, between the Daily thoracic impedance value and the Reference
thoracic impedance value.
The Thoracic Impedance trend plots the Daily and Reference Impedance values.
6.7.5.2 Viewing OptiVol observations
Select Data icon
⇒ Quick Look II
A clinical status observation appears on the Quick Look II screen and on the Heart Failure
Management Report when the OptiVol 2.0 Fluid Index has reached or exceeded the OptiVol
Threshold since the last session. If the OptiVol 2.0 Fluid Index is still greater than 0, the
observation displays the date of the first day that the OptiVol 2.0 Fluid Index was equal to
or greater than the threshold, and “ongoing”. If the OptiVol 2.0 Fluid Index has since reset
to 0, the observation displays the date of the first day that the OptiVol 2.0 Fluid Index was
equal to or greater than the threshold, and the date that the OptiVol 2.0 Fluid Index reset to
0.