Figure 1. System components
VVI
Medtronic CareLink Programmer and
Analyzer
Conexus Activator
Medtronic CareLink Monitor
Implantable device
system
InCheck Patient
Assistant
AF?
Clinic
Home
Programmers and software – The Medtronic CareLink Model 2090 Programmer and
software are used to program this device. The Medtronic CareLink Model 2090 Programmer
with Conexus wireless telemetry is designed to provide clinicians and patients with an easy
and efficient implant, follow-up, and monitoring experience. Conexus wireless telemetry
eliminates the need to have a programming head placed over the implanted device for the
duration of a programming or monitoring session. The system uses radio frequency (RF)
telemetry for wireless communication between the implanted device and programmer in the
hospital or clinic. Conexus telemetry operates within the Medical Implant Communications
Service (MICS) Band, which is the only band designated for implantable medical devices.
Using the MICS Band prevents interference with home electronics such as microwaves,
cell phones, and baby monitors.
To turn on Conexus telemetry in an implanted device, you must use the Conexus Activator
or the programming head. If you do not use the Conexus Activator or if you are using a
programmer with nonwireless telemetry, you will need to use the programming head to both
initiate and conduct communications with the device in the clinic.
Medtronic
PROTECTA™ XT DR D314DRG
20
Clinician Manual