or permanently elevate pacing thresholds or temporarily or permanently damage the
myocardium. If external defibrillation or cardioversion are required, consider the following
precautions:
●
Use the lowest clinically appropriate energy.
●
Position the patches or paddles a minimum of 15 cm (6 in) away from the device.
●
Position the patches or paddles perpendicular to the device and leads.
●
If an external defibrillation or cardioversion is delivered within 15 cm (6 in) of the device,
use a Medtronic programmer to evaluate the device and lead system.
Hyperbaric therapy (including hyperbaric oxygen therapy, or HBOT) – Hyperbaric
therapy is the medical use of air or 100% oxygen at a higher pressure than atmospheric
pressure. Hyperbaric therapies with pressures exceeding 2.5 ATA (approximately 15 m (50
ft) of seawater) may affect device function or cause device damage. To avoid or mitigate
risks, do not expose implanted devices to pressures exceeding 2.5 ATA.
Lithotripsy – Lithotripsy is a medical procedure that uses mechanical shock waves to break
up kidney or gallbladder stones. If the device is at the focal point of the lithotripter beam,
lithotripsy may permanently damage the device. If lithotripsy is required, keep the focal point
of the lithotripter beam a minimum of 2.5 cm (1 in) away from the device. To avoid or mitigate
the effects of oversensing, consider the following precautions:
●
Suspend tachyarrhythmia detection by using a magnet or a programmer. After
completing lithotripsy treatment, remove the magnet or restore device parameters.
●
If appropriate for the patient, program the device to an asynchronous pacing mode (for
example, DOO). After completing lithotripsy treatment, restore device parameters.
Magnetic resonance imaging (MRI) – An MRI is a type of medical imaging that uses
magnetic fields to create an internal view of the body. Do not conduct MRI scans on patients
who have this device or lead implanted. MRI scans may result in serious injury, induction
of tachyarrhythmias, or implanted system malfunction or damage.
Radiotherapy – Radiotherapy is a cancer treatment that uses radiation to control cell
growth. When performing radiotherapy, take precautions to avoid oversensing, device
damage, and device operational errors, as described in the following sections:
●
Oversensing – If the patient undergoes radiotherapy treatment and the average dose
rate at the device exceeds 1 cGy/min, the device may inappropriately sense direct or
scattered radiation as cardiac activity for the duration of the procedure. To avoid or
mitigate the effects of oversensing, consider these precautions:
– Suspend tachyarrhythmia detection by using a magnet or a programmer. After
completing radiotherapy treatment, remove the magnet or restore device
parameters.