When evaluating an ECG strip, you will notice that the VSP pulse appears shortly after a
ventricular sense and usually has a shorter AV interval. The “VP” annotation in the Marker
Channel usually does not appear on a printed real-time ECG strip due to the limited space
after the “VS” annotation. Both the “VP” and the “VS” annotations appear in the Live Rhythm
Monitor, on frozen strips, and on printed frozen strips.
7.15 Preventing rapid ventricular pacing during atrial
tachyarrhythmias
An atrial tachyarrhythmia may result in a rapid ventricular pacing rate when the device is
operating in the DDDR or DDD mode. The implanted device should be capable of
withholding atrial tracking during periods of atrial tachyarrhythmia, while tracking the normal
sinus rate.
7.15.1 System solution: Mode Switch
The Mode Switch feature switches the device pacing mode to a nontracking mode upon
detection of an atrial tachyarrhythmia and restores the programmed pacing mode when the
atrial tachyarrhythmia ends. By operating in a nontracking mode, the device prevents rapid
ventricular pacing that may result from a high atrial rate.
7.15.2 Operation of Mode Switch
Figure 115. Overview of Mode Switch operation
Normal Sinus
AT/AF
DDD(R)
DDIR
Mode Switch is available when the pacing mode is programmed to DDDR, DDD, or an MVP
mode (AAIR<=>DDDR or AAI<=>DDD). Mode Switch functions when the device is
operating in the DDDR or DDD mode.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
273