determine the new amplitude setting. The device then adjusts the current Atrial Amplitude
toward this target. The device reduces the amplitude by 0.25 V every other day until it
reaches the target amplitude. If the operating amplitude is below the target, the device
adjusts it to the target immediately. The lower limit is set by the programmed Atrial Minimum
Adapted Amplitude. If the operating pulse width has a value different from 0.4 ms, the device
adjusts it to that value.
Upper limit for adjustments – The device adjusts the Atrial Amplitude to 5.0 V and the
Atrial Pulse Width to 1.0 ms if the amplitude threshold is greater than 2.5 V or the target
amplitude is greater than 5.0 V.
7.5.3.4 Stopping an atrial pacing threshold search in progress
The device stops a pacing threshold search immediately if there are sudden changes in the
patient’s heart rate or if other device features take precedence over the search.
When a pacing threshold search cannot be completed, the device automatically schedules
another search within 30 min. If 5 more search attempts are stopped during a day, the pacing
threshold test is suspended until the following day. Whenever this happens, a device check
occurs again, and the process is repeated. The reasons for stopping a pacing threshold
search are noted in the Capture Threshold trends diagnostic (see Section 7.5.7).
7.5.4 Operation of Right Ventricular Capture Management
Right Ventricular Capture Management (RVCM) is available when the pacing mode is
programmed to DDDR, DDD, DDIR, DDI, VVIR, VVI, or an MVP mode (AAIR<=>DDDR or
AAI<=>DDD), and functions when the device is operating in one of these modes. If RVCM
is programmed to the Monitor or Adaptive setting, the device conducts a pacing threshold
search to determine the RV pacing threshold. If RVCM is programmed to the Adaptive
setting, the device uses the RV pacing threshold to define a target amplitude and adjusts
the pacing amplitude toward the target amplitude. The target amplitude is based on the
programmed settings for the RV Amplitude Safety Margin and the RV Minimum Adapted
Amplitude parameters.
Note: In the event of partial or complete lead dislodgment, RVCM may not prevent loss of
capture.
Note: If the battery reaches the Recommended Replacement Time (RRT), the device aborts
RVCM. No additional RV pacing threshold searches will be conducted.
Medtronic
PROTECTA™ XT DR D314DRG
244
Clinician Manual