10.7.2.2 How to induce VF with T-Shock
Figure 220. T-Shock induction screen
1. Select Tests > EP Study.
2. Select T-Shock from the list of inductions and therapies.
3. If you want to treat the induced episode with a manual therapy, select [Suspend] to
prevent automatic detection.
Note: The [Suspend] button is located at the top of the screen and is not shown in
4. Select the Resume at DELIVER check box for automatic detection and therapy, or
clear the check box for manual therapy.
Note: During a wireless telemetry session, you cannot deliver a T-Shock induction
when there is a magnet or programming head over the device and the Resume at
DELIVER check box is selected. If an error message appears, remove the magnet or
programming head, or clear the Resume at DELIVER check box.
5. Accept the displayed test values or select new test values.
6. To view and adjust VF detection and therapy parameters, select [Adjust
Permanent…].
7. Select the Enable check box.
8. Select [DELIVER T-Shock].
Note: If the energy on the capacitors is higher than the energy level you selected, the
programmer displays a warning when you select [DELIVER T-Shock]. To clear this
warning select either [DUMP] or [CANCEL].
9. If necessary, select [ABORT] to abort the induction or any therapy in progress.
10. Select [Retrieve Data…] and [Print…] to review and print patient data.