Figure 102. Rate Drop Response Rate and Time
Hea
rt R
ate
Time
Sinus rate
Intervention
Duration (2 min)
Step-down
(approx 7 min)
Detection
As shown in Figure 102, Rate Drop Response typically operates over several minutes, and
most of this time involves the step-down phase.
Rate Drop Response is available when the pacing mode is programmed to DDD, DDI, or
AAI<=>DDD (MVP mode). Rate Drop Response functions when the device is operating in
the DDD or DDI mode. For the MVP mode, the device operates in DDD mode during Rate
Drop Response interventions. Rate Drop Response does not operate during
tachyarrhythmias, Mode Switch episodes, and Capture Management pacing threshold
searches.
7.8.2.1 Detection
Rate Drop Response provides 2 methods for detecting significant rate drops:
●
Drop Detection
●
Low Rate Detection
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
257