Figure 164. Wavelet performance with varying Match Threshold values
40%
97%
Match Threshold value
70% (nominal)
Increased specificity
Decreased sensitivity
Decreased specificity
Increased sensitivity
1 With a decrease in the Match Threshold value, the device is more likely to withhold detection
of rapidly conducted SVTs (increased specificity) but is less likely to detect true VT (decreased
sensitivity).
2 With an increase in the Match Threshold value, the device is less likely to withhold detection of
rapidly conducted SVTs (decreased specificity) but is more likely to detect true VT (increased
sensitivity).
Note: In the Marquis VR Single Chamber ICD clinical study, the nominal Match Threshold
of 70% was used in 99% of episodes in 99% of the patients studied.
Missing template – If Wavelet is enabled but no template is available, detection will occur
as if Wavelet were disabled until a new template is stored.
8.4.4 Programming Wavelet
8.4.4.1 Programming Wavelet
Select Params icon
⇒ Detection (V.)…
⇒ Wavelet…
▷ Wavelet <On, Monitor>
▷ Match Threshold
▷ Auto Collection
Medtronic
PROTECTA™ XT DR D314DRG
330
Clinician Manual