![Medtronic PROTECTA XT DR D314DRG Скачать руководство пользователя страница 171](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148171.webp)
2 Use the Plot check boxes to display ventricular intervals, atrial intervals, or both.
3 This portion of the display shows the programmed detection intervals.
Note: The device may truncate data storage during an episode to conserve device memory.
If so, the programmer displays time labels on the horizontal axis of the interval plot following
the truncation as asterisks (*).
6.8.3.2 Viewing the episode EGM
When you select an episode from the episode log and then select the EGM option, the
programmer displays the stored EGM data for the episode.
Figure 48. Episode EGM
1 The Marker Channel displays the annotated atrial and ventricular events, leading up to
detection.
2 The Decision Channel displays an annotation conveying the type of episode detected (here VT
Monitor). The EGM display must be maximized to display the Decision Channel annotations.
3 Use the horizontal scroll bar at the bottom of the screen to view all of the episode EGM data.
4 Use this button to select an option for displaying one of the atrial intervals. The EGM display
must be maximized to select the atrial interval display options.
5 This annotation provides the amount of time EGM recording was suspended to conserve
storage space.