Figure 58. Battery and Lead Measurements screen
1 Battery voltage and replacement indicator
information
2 Capacitor charging information
3 Sensing Integrity Counter
4 Result of the last Atrial Lead Position Check
5 Most recent lead impedance
measurements
6 Most recent daily automatic sensing
amplitude measurements
7 Information about the last delivered high
voltage therapy
8 Select [Print…] to print a Battery and Lead
Measurements Report
6.13.1.2 Battery voltage and replacement indicators
The device automatically measures the battery voltage when telemetry is initiated at the
start of a session, when a lead impedance test is performed, and every day at 2:15 AM as
part of the automatic daily measurements. The most recent battery voltage measurement
is displayed on the Battery and Lead Measurements screen.
Note: You may see a temporary drop in the displayed battery voltage if high-voltage
charging has occurred within the past 24 hours.
If 3 consecutive automatic daily battery voltage measurements are less than or equal to the
Recommended Replacement Time (RRT) value, the programmer displays the RRT symbol
and the date when the battery reached RRT. If the programmer displays the RRT symbol,
contact your Medtronic representative and schedule an appointment to replace the device.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
187