The daily automatic sensing amplitude measurements are displayed on the Lead Trends
screen, which plots the data as a graph. The graph displays up to 15 of the most recent
measurements and up to 80 weekly summary measurements (showing minimum,
maximum, and average values for each week). Significant or sudden changes in sensing
amplitude may indicate a problem with a lead.
Note: The sensing amplitude trend data is intended to show changes in sensing amplitude
measurements that may be used to assess lead integrity. The adequacy of the ventricular
sensing safety margin cannot be determined by the R-wave trend measurement, and should
be based on VF induction testing.
6.13.3.1 How to view sensing amplitude trends
Select Data icon
⇒ Device/Lead Diagnostics
⇒ P/R Wave Amplitude Trends
Figure 60. Lead Trends screen showing the R-wave amplitude trend
1 Selected measurement trend
2 Selected amplitude measurement type
3 Weekly minimum, maximum, and average
values
4 Most recently measured values
5 Last automatic daily measurement
6 Select [Print…] to print a Lead Trends
Report
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
191