8.8.6.5 Medtronic CareAlert Events
Select Data icon
⇒ Alert Events
Check the CareAlert Events list to verify that there is an RV lead noise warning.
8.9 Discriminating T-wave oversensing from VT/VF
T-wave oversensing occurs when a device senses T-waves in addition to R-waves. This
double-counting of ventricular events may push the ventricular rate into the VT/VF zone,
lead to incorrect detection of VT or VF, and deliver inappropriate therapy.
8.9.1 System solution: TWave Discrimination
The TWave Discrimination feature withholds VT/VF detection when a fast ventricular rate
is detected because of oversensed T-waves. It analyzes differences in amplitude, rate, and
pattern to differentiate R-waves from T-waves. It then applies additional criteria to
distinguish R and T patterns from VT/VF. This reduces the potential to deliver inappropriate
therapy for high rates that are attributable to T-wave oversensing.
8.9.2 Operation of TWave Discrimination
When VT or VF is suspected, TWave Discrimination applies amplitude, rate, and pattern
criteria to determine if both R-waves and T-waves are being sensed. If both are sensed, VT
or VF detection is withheld. If only R-waves are sensed, VT or VF detection occurs.
Medtronic
PROTECTA™ XT DR D314DRG
350
Clinician Manual