% of Time – This section shows the patient’s conduction status as the percentage of the
total time that the device paced or sensed during the collection period. The percentages
are calculated from the daily counts of AS-VS, AS-VP, AP-VS, and AP-VP event sequences.
Atrial rate histogram – The atrial rate histogram shows the rate distribution of atrial sensed
and paced events (including sensed events that occur during the refractory period). The
histogram also indicates if the percentage of atrial senses that may be due to far-field
R-wave (FFRW) sensing is 2% or greater. If so, the percentage is reported within one of two
ranges: 2% to 5% of AS may be due to FFRW or > 5% of AS may be due to FFRW. Far-field
R-wave sensing may be suspected if the intervals between atrial sensed events are
irregular.
Ventricular rate histogram – The ventricular rate histogram shows the rate distribution of
ventricular sensed and paced events.
Ventricular rate during AT/AF histogram – The ventricular rate during AT/AF histogram
shows ventricular sensed and paced events that occurred during detected atrial
arrhythmias, and the total time in AT/AF
. This histogram may be used to monitor the
effectiveness of ventricular rate control therapy and drug titration.
6.13 Viewing detailed device and lead performance data
The device automatically measures and records device and lead performance data every
day. Detailed views of this data are available from the Battery and Lead Measurements
screen and the Lead Trends screen.
6.13.1 Viewing battery and lead measurement data
The Battery and Lead Measurements screen displays the most recent values for key
measures of device and lead performance. These may include automatically measured
values or those measured during manual system tests.
6.13.1.1 How to view battery and lead measurement data
Select Data icon
⇒ Device/Lead Diagnostics
⇒ Battery and Lead Measurements
7
“Detecting atrial tachyarrhythmias”, page 295.
Medtronic
PROTECTA™ XT DR D314DRG
186
Clinician Manual