Chapter 8 Replacement Procedures
8-131
8-8-3-6 JVC VCR Installation Procedure
(cont’d)
9.) Bundle all the VCR cables together into a single loom and bind together with cable ties.
10.)With the control console still at the
maximum
height, route the cable loom down to the
left
rear
connectors panel (at the rear of the
Vivid 3
unit).
11.)Fasten the loom to the Vivid 3N bottom cover using the square adhesive cable fastener (supplied).
12.)Connect the RS-232C cable, the Audio and S-Video IN and OUT cables, and the Power cable to
the appropriate sockets on the Vivid 3N rear connectors panel - see
Figure 8-115 "Vivid 3N Left
Rear Panel Connectors" on page 8-132
.
Figure 8-114 JVC VCR Installed in Compartment
JVC VCR
Fix Adhesive Velcro Strip Here