GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
3-66
Section 3-9 - Connectivity Setup
3-9-7-1 Validating Communication with the EchoPAC Server
(cont’d)
7) If both connections were established, click
OK
to exit the configuration procedure. If not, review the
entire procedure to locate the error.
NOTE:
IMPORTANT INFORMATION ABOUT DATAFLOWS
:
The system’s default dataflow is
Local Archive - Int. HD
. In this configuration, the Patient List
is in the local archive and the images are stored on the internal hard drive.
In a dataflow configuration such as
Int. HD - MOD 525 (DICOM)
, images are stored both on the
internal hard drive and on media (MOD). The name DICOM is shown in parenthesis, signifying
that on the MOD, the images are in DICOM format.
No Archive
is a dataflow that has no output source, for use in emergencies.
CAUTION: DO NOT change the name of an existing dataflow or attempt to add new services to
a dataflow!
In the event that any changes or additions are required, contact the On-line Center.
3-9-8
Validating EchoPAC or Image Vault Operation
1) Press the
Archive
button on the control console. The
Patient List
window is displayed, as shown in
Figure 3-43 "Patient List Window" on page 3-60
.
2) Select the
Remote Archive-Remote Storage
option from the dropdown list at the bottom of the
window. If the system will not accept the dataflow, or displays an error message, there is an error
in dataflow communication.
3) Begin an exam for a new patient and store some images. The storage process will take longer as
each loop is now stored in both raw data and DICOM format.
4) Begin an exam for a new patient. Once the second patient is active, the first patient’s exam is closed
and sent to the spooler to be sent over the network.
5) Press
<Ctrl+S>
to view the status of the spooler, as described on page
3-61
.
Содержание Vivid 3N Pro Series
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL ii iii...
Страница 24: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL xxii...
Страница 597: ......