Chapter 7 Diagnostics/Troubleshooting
7-7
7-3-1-1 Auto Sequence
(cont’d)
Auto Sequence commences. While the sequence of automatic testing proceeds, the name of the
test currently in progress is displayed in the
Test Name
field and the current
Loop #
and
Step #
are
shown below. In addition, a flashing colored light below the
Passed
and
Failed
displays, indicates
the current test status, as follows:
•
Green
= test
passed
•
Yellow
= test in progress
•
Red
= test failed
5.) At any stage, trackball to the
Pause
button (under
Execute Sequence)
and press
Select
, if required.
6.) To resume testing (from the point where you paused), trackball to the
Start
button and press
Select.
7.) At any stage during the testing sequence, trackball to the
View Report
button and press
Select
to
review the current Auto Sequence Test Report.
8.) When Auto Sequence is complete, trackball to the
View Report
button and press
Select
to review
the final Auto Sequence Test Report. This will now list all tests performed and show details of which
tests passed or failed.
9.) Trackball to the
Exit
button and press
Select
to close the
Automatic
FE Testing
dialog box.
NOTE:
During Auto Sequence, the system will automatically re-boot at least twice. This is normal and should
not be cause for concern.