Chapter 7 Diagnostics/Troubleshooting
7-81
7-4-6-1 MOD Diagnostic Test
(cont’d)
3) Trackball to the
MO
radio button and press
Select
.
4) Trackball to the
IsMediaAccess
button an press
Select
. The message
Media in drive G
is
accessible is displayed.
5) Trackball to the
Eject
button and press
Select
. The message
Media is outside
is displayed. Verify
that the media has actually been ejected.
6) Trackball to the
Load
button and press
Select
. The message
Media is inside
is displayed. Verify
that the media is actually inside.
7) Trackball to the
Format
button and press
Select
. The message
Media in drive G was formatted
successfully
is displayed.
8) Trackball to the
GetVolInfo
button and press
Select
. The message
Media in drive G has
following properties: Media Label = MY_LABEL, Media File-System Name = FAT
is displayed.
9) Trackball to the
GetBtPSector
button and press
Select
.
The message The Bytes Per Sector
value of media in drive G is 512
is displayed.
10) Trackball to the
SetVolLabel1
button and press
Select
.The message
Media Label set to
TEST_LABEL1
is displayed.