Chapter 7 Diagnostics/Troubleshooting
7-39
7-3-8-4 Receive Test
(cont’d)
The
Report
dialog box contains the following information:
•
Test Result Message:
Displays the test result.
•
TR4 Modules:
Displays a graphic representation of the front board (four groups of TR4 sockets).
Each group consists of four TR4 sockets, each socket containing four vectors (channels), as
follows:
-
Each vector is represented by a button (highlighted in green) that displays the number of the
socket to which the TR4 is connected.
-
Each button is highlighted yellow when selected.
-
Each button is highlighted red if errors were detected for one or more channels.
•
Checkboxes:
Enable you to manipulate bad channels:
-
Show Only Bad Channels:
Select this checkbox to display only bad channels in the
Display
area.
-
Mark bad channel:
Select this checkbox to manually mark the currently selected channel as
bad.
Figure 7-21 Report Dialog Box - Receive Test Low Frequency