GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
8-12
Section 8-2 - Cover Replacement Procedures
8-2-6-3 Top Cover Lower Section Removal Procedure
(cont’d)
13) Remove the three screws securing the remaining section of the gas spring cover, shown below, and
lower it into the collar of the top cover.
Figure 8-14 Gas Spring Cover Screws
CAUTION: The remaining section of the gas spring cover cannot be removed at this stage. Do
not attempt to remove it.
14) Remove the three screws located at the front edge of the top cover, and allow it to sink down into
the top cover collar.
15) Slightly lift the top cover, together with the remaining section of the gas spring cover, and turn them
90 degrees
clockwise
.
16) Keep the opened gap of the gas spring cover parallel to the opened gap of the top cover’s collar,
and pull them out together in a
clockwise
direction.
17) Remove the gas spring cover section from the top cover.
8-2-6-4
Top Cover Lower Section Installation Procedure
1) Return the gas spring cover section back to its original position in the top cover.
2) Align the opened gap of the gas spring cover to be in parallel with the opened gap of the top cover
collar.
3) Hold the top cover and the gas spring cover section so that they are at a 90-degree angle to the
front of the unit, then insert them from the
left
side of the unit through their opened gap.
4) Insert and screw the three screws located at the top edge of the top cover, shown in
Figure 8-5 "Top
Cover Screws (Lower Section)" on page 8-6
.
5) Lift the gas cover spring cover from the top cover and fasten it with the three screws in its original
location.
6) Return the rear section of the gas spring cover to its original position and fasten it with the two
screws shown in
Figure 8-14 "Gas Spring Cover Screws" on page 8-12
.
7) Gently fit the left and right connector panels back to their original position on the top cover and
fasten each with its four screws.
8) Mount the small section of the top cover collar, closing the open gap at the rear of the top cover,
and fasten it to the top cover with the two screws shown in
Figure 8-11 "Top Cover Collar" on page
8-10
.
Содержание Vivid 3N Pro Series
Страница 2: ......
Страница 5: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL ii iii...
Страница 24: ...GE MEDICAL SYSTEMS DIRECTION FK091075 REVISION 04 VIVID 3N PRO EXPERT SERVICE MANUAL xxii...
Страница 597: ......