GE M
EDICAL
S
YSTEMS
D
IRECTION
FK091075, R
EVISION
04
V
IVID
3N P
RO
/E
XPERT
S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8-31
8-3-2
Vivid 3N 17" Monitor Replacement - Procedure 1
NOTE:
This section describes the procedure for replacing a 17" monitor based on a Samsung 17" monitor type.
8-3-2-1
Tools
Use the appropriate Phillips-type screw drivers as indicated in the Vivid 3N
monitor replacement
procedures.
8-3-2-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
Chapter 3 -
Installation
.
8-3-2-3
Vivid 3 17" Monitor Removal - Procedure 1
1.) Turn OFF the Vivid 3N main circuit breaker, located at the rear of the machine (lower end).
2.) Disconnect the main AC power cable from the wall outlet.
3.) Unplug the VGA cable and AC power cable from the rear control console.
4.) Remove the four screws that fasten the plastic cover to the base assembly (two from each side).
Remove the plastic cover.
5.) Loosen the securing screw located at the rear of the monitor base - see
Figure 8-32 "Releasing
Screw at Base of 17" Monitor" on page 8-31
.
Once this screw is loosened, the monitor may be fully rotated and lifted upwards, away from the
machine.
6.) Remove the monitor by pulling it upwards.
While pulling the monitor upwards, swiveling it slightly to the
right
and
left
will ease removal.
NOTE:
When removing the monitor, make sure that the Teflon washer remains in place on top
of the monitor mounting base.
You are now ready to install a new monitor, as described in the following procedure.
Figure 8-32 Releasing Screw at Base of 17" Monitor