Chapter 8 Replacement Procedures
8-23
8-2-12
Right and Left Probe Holders Replacement Procedure
8-2-12-1
Tools
Use the appropriate Phillips-type screw drivers and an 8mm wrench, to replace the
right
and
left
probe
holders.
8-2-12-2
Preparation
Shut down the Vivid 3N Pro/Expert ultrasound unit, as described in
Chapter 3 -
Installation
.
8-2-12-3
Right and Left Probe Holders Removal Procedure
1) Remove the bottom keyboard cover, as described in
8-2-8-3 "Bottom Keyboard Cover Removal
Procedure" on page 8-14
.
2) Remove the two 8mm securing nuts (located below the keyboard) that fasten each of the probe
holders to the metal frame, as shown below:
Figure 8-24 Probe Holder Nuts
3) Remove the probe holders.
8-2-12-4
Right and Left Probe Holders Installation Procedure
1) Return the probe holders to their original position on either side of the keyboard.
2) Fasten each probe holder beneath the keyboard with its two 8 mm securing nuts.
3) Return the bottom keyboard cover to its original position, as described in
8-2-8-4 "Bottom Keyboard
Cover Installation Procedure" on page 8-15
.